9 Programming the Instrument
9.4 Status Register System & STATus Subsystem
Questionable Temperature Event Query
Returns the decimal value of the sum of the bits in the Questionable Temperature
Event register.
NOTE
The register requires that the associated PTRansition or NTRansition filters be
set before a condition register bit can set a bit in the event register. The data in
this register is latched until it is queried. Once queried, the register is cleared.
Remote Command
:STATus:QUEStionable:TEMPerature[:EVENt]?
Example
:STAT:QUES:TEMP?
Preset 0
Status Bits/OPC depend-
encies
Sequential command
Questionable Temperature Negative Transition
Determines which bits in the "Questionable Temperature Condition" on page 1124
Register will set bits in the "Questionable Temperature Event Query" on page 1125
register, when the condition register bit has a negative transition (1 to 0).
The variable <integer> is the sum of the decimal values of the bits that you want to
enable.
Remote
Command
:STATus:QUEStionable:TEMPerature:NTRansition <integer>
:STATus:QUEStionable:TEMPerature:NTRansition?
Example
:STAT:QUES:TEMP:NTR 2
Instrument Overheated being cleared will be reported to the Temperature Summary of the Status
Questionable register
Preset 0
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
Questionable Temperature Positive Transition
Determines which bits in the "Questionable Temperature Condition" on page 1124
Register will set bits in the "Questionable Temperature Event Query" on page 1125
register, when the condition register bit has a positive transition (0 to 1).