9 Programming the Instrument
9.4 Status Register System & STATus Subsystem
Questionable Integrity Output Positive Transition
Determines which bits in the "Questionable Integrity Output Condition" on page
1150 register will set the corresponding bit in the "Questionable Integrity Output
Event Query" on page 1151 register when the condition register bit has a positive
transition (0 to 1).
The variable <integer> is the sum of the decimal values of the bits that you want to
enable.
Remote
Command
:STATus:QUEStionable:INTegrity:OUTPut:PTRansition <integer>
:STATus:QUEStionable:INTegrity:OUTPut:PTRansition?
Example
:STAT:QUES:INT:OUTP:PTR 1
Oversweep set is reported to the Integrity Summary of the Status Questionable register
Preset 32767
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
9.4.6.16 Questionable Integrity Uncalibrated Register
Monitors conditions that occur because a measurement may not be able to return
an accurate or valid result due to a mismatch between instrument settings and the
signal, placing the instrument in an uncalibrated state for that signal. It summarizes
them in bit 3 of the "Questionable Integrity Register" on page 1142.
Bit Condition Operation
0 Meas Uncal A Meas Uncal warning is being displayed; generally this means the
sweep time must be reduced or the RBW increased
1 Signal ID on In external mixing, the Sig ID function is on, which will impact the
IQ Analyzer Mode User's &Programmer's Reference 1152