LB 9507 Chapter 7: Quantitative Measurements (LIA...)
7-7
L12
08: AUTOMATIC BKG SUBTRACTION YES
YES NO LAST DISPLAY
Here you can select whether the background effect is to be
subtracted automatically from the measured results. Remember
that the statistical significance of the measured values is
dependent upon the measuring time.
L13
08: MAX. BACKGROUND [RLU/s] 50
(0 = NONE) LAST DISPLAY
Here can you define the limit value for the background. If this
threshold is exceeded, you will be alerted accordingly on the
display and the printout. Entering “0” means "no background
threshold“. As soon as you confirm the entry with <enter> you
move to the next display.
L14
08: DELAY LAST INJ./MEAS. [s] 1.0
(0.0 - 300.0) LAST DISPLAY
Enter the delay between the last injection and the start of the
measurement in a range from 0.0 to 300 seconds and press
<enter>.
L15
08: MEASURING TIME [s] 2.0
(0.1 - 200.0) LAST DISPLAY
Enter the desired measuring time (0.1 to 200.0 seconds) for each
tube and press <enter>.
L16
08: CONC. UNITS NG/ML
LAST DISPLAY
Enter the units of concentration, e.g. ng/mL, and press <enter>.
You can enter a maximum of 6 characters.