LB 9507 Chapter 7: Quantitative Measurements (LIA...)
7-11
L35
08: TRANSF. 0=LOGIT/LG 1=LG/LG 0
LAST DISPLAY
Choose whether to perform a logit-log or log-log transformation
prior to the spline fit. Use Logit-log for standard curves where
saturation is expected in the upper RLU range (usually
preferable). You can select log-log instead (see Appendix A-1 for
more information).
L36
08: REPL. OF TOTAL (0-10) 0
LAST DISPLAY
Enter the replicate number for totals (total reagent). Not relevant
to assay evaluation.
L37
08: REPL. OF NSB (0-10) 0
LAST DISPLAY
Enter the replicate number for NSB and press <enter>. The value
measured with these samples is subtracted from the standards,
controls and patient samples (see A-3).
L38
06: INPUT ALL CORRECT ?
YES NO LAST DISPLAY
If you select <YES> the entered parameters are stored and the
program returns to the main menu (L1). <NO> takes you back to
the 1st display of this protocol (L5). Entries and changes are only
saved if you exit the routine by entering <YES> at L23.
Note: The protocol is not stored if you exit this protocol sequence
via <exit>!