Chapter 7: Quantitative Measurements (LIA...) LB 9507
7-10
L29
08: REPL. OF SAMPLES (1-10) 1
LAST DISPLAY
Enter the replicate number of patient samples (1 - 10).
L30
USE COMPLETE STANDARD SERIES ? YES
YES NO LAST DISPLAY
Select <YES> to use a complete standard series. If you select
<NO> the protocol setup is finished. Now you have to enter the
values of the master curve in the submenu <MASTER CURVE>.
L31
08: NO. OF STANDARDS (4-10) 4
LAST DISPLAY
Enter the number of standards and press <enter>.
L32
08: CONC. OF STANDARD 1 0.000
LAST DISPLAY
Enter the concentration of standard no. 1.
L33
08: CONC. OF STANDARD 2 0.000
LAST DISPLAY
Enter the concentration of standard no. 2 and then continue for all
standards specified, pressing <enter> to confirm. The
concentration values must be entered in order (1 to 10), otherwise
an error message will appear.
L34
08: PR. TYPE 0=ILMA 1=LIA/SPALT 0
LAST DISPLAY
Select the curve type for an immunoluminometric (ILMA) or
competitive (LIA/SPALT) assay by entering “0” or “1” and pressing
<enter>.