LB 9507 Chapter 7: Quantitative Measurements (LIA...)
7-9
L23
08: CONTROL 2 LOW LIMIT 0.00
LAST DISPLAY
Enter the lower limit for the 2nd control sample and press
<enter>.
L24
08: CONTROL 2 HIGH LIMIT 14.00
LAST DISPLAY
Enter the upper limit for the 2nd control sample and press
<enter>.
L25
08: EXP. RANGE LOW LIMIT 0.000
LAST DISPLAY
Enter the lower limit of the expected range (normal range) and
press <enter>.
L26
08: EXP. RANGE HIGH LIMIT 0.000
LAST DISPLAY
Enter the upper limit of the expected range (normal range) and
press <enter>.
L27
08: REPL. OF CAL./STD. (1-10) 1
LAST DISPLAY
Enter the replicate number of the standards (1 - 10).
L28
08: REPL. OF CONTROLS (1-10) 1
LAST DISPLAY
Enter the replicate number of the control samples (1 - 10),
assuming you have selected controls in display L18.