Chapter 7: Quantitative Measurements (LIA...) LB 9507
7-20
Start the measurement for the 2nd tube. The measurement
procedure will be performed provided the background limit is not
exceeded. The steps which the instrument goes through are
displayed, and the measurement parameters from the active
protocol are printed out at the same time.
BACKGROUND... INJECTION... MEASURING...
L65
SAMPLE-RLU 133
REMOVE TUBE
Remove the sample tube.
L66
NSB REPL. # 1 TUBE # 3
INSERT TUBE ...
Load the 3rd sample tube. In our example this is the 1st NSB tube
(it is the third tube of this measurement series).
L66a
NSB REPL. # 1 TUBE # 3
START
Press the <start> button.
Following this measurement, Lumat waits for the other NSB tubes
defined in the protocol. The following then appears:
L67
STD # 1 REPL. # 1 TUBE # 6
INSERT TUBE ...
Lumat waits for you to load the 1st standard tube. Measurement
begins once you have positioned the tube and pressed the
<start> button.
BACKGROUND... INJECTION... MEASUREMENT...