LB 9507 Chapter 8: T-Uptake/FTI-Measurements
8-7
T12
09: AUTOMATIC BKG SUBTRACTION NO
YES NO LAST DISPLAY
Here you can select whether the background effect is to be
subtracted automatically from the measured results. Remember
that the statistical significance of the measured values is
dependent upon the measuring time.
T13
09: MAX. BACKGROUND [RLU/s] 50
(0 = NONE) LAST DISPLAY
Here can you define the limit value for the background. If this
threshold is exceeded, you will be alerted accordingly on the
display and the printout (BKG).
Entering “0” means “no background threshold“.
As soon as you confirm the entry with <enter> you move to the
next display.
T14
09: DELAY LAST INJ./MEAS. [s] 1.0
(0.0 - 300 s) LAST DISPLAY
Enter the delay between the last injection and the start of the
measurement in a range from 0.0 to 300 seconds and press
<enter>.
T15
09: MEASURING TIME [s] 2.0
(0.1 - 200.0) LAST DISPLAY
Enter the desired measuring time (0.1 to 200.0 seconds) for each
tube and press <enter>.
T16
09: CONC. UNITS NG/ML
LAST DISPLAY
Enter the units of measure for concentrations, e.g. ng/mL, and
press <enter>. You can enter up to 6 characters.