EasyManua.ls Logo

LUMAT LB 9507 - Page 128

LUMAT LB 9507
222 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 8: T-Uptake/FTI-Measurements LB 9507
8-8
T17
09: NUMBER OF DECIMAL PLACES (0-3) 2
LAST DISPLAY
Enter the desired number of decimal places for the calculated
concentration in all printouts and press <enter>.
T18
09: NO. OF CONTROLS (0-10) 2
LAST DISPLAY
Enter the number of control samples used for the measurements
(0 to 10) and press <enter>.
T19
09: CONTROL 1 ID.
LAST DISPLAY
Enter an alphanumeric name for the 1st control sample (so you
can remember it later when performing the measurement, since
the name is displayed on the LCD) and press <enter>.
T20
09: CONTROL 2 ID.
LAST DISPLAY
Enter a name for the 2nd control sample, and subsequently all the
other controls you have defined.
T21
09: CONTROL 1 LOW LIMIT 0.000
LAST DISPLAY
Enter the lower limit for the 1st control (digits corresponding to the
format defined in T17) and press <enter>.
T22
09: CONTROL 1 HIGH LIMIT 0.000
LAST DISPLAY
Enter the upper limit for the 1st control sample and press <enter>.

Table of Contents