Chapter 10: Kinetic Measurements LB 9507
10-8
Enter the delay between the last injection and the start of the
measurement in the range from 0.0 to 300 seconds and press
<enter>.
K13
06: MEASURING TIME [s] 2.0
(0.1 - 200.0) LAST DISPLAY
Enter the desired measuring time (0.1 to 200.0 seconds) for each
data point and press <enter>.
K14
06: Meas.Time Total [s] 2
(2 - 5000) LAST DISPLAY
Enter the total measurement time which indicates the time that the
tube is in measurement position and press <enter>. You can
measure in a range from 2 up to 5000 seconds.
K15
06:Delay Time Betw. M. [s] 2
(2 300.0) LAST DISPLAY
Enter a delay time between the single measurements of one
sample. This setting allows you to set up an individual interval
time scale for each measurement
K16
06: INPUT ALL CORRECT ? YES
YES NO LAST DISPLAY
If you select <YES> the entered parameters are stored and the
program returns to the main menu (K1).
<NO> takes you back to the 1st display of this protocol (K5).
Note:
The protocol is not stored if you exit this protocol sequence via
<exit>!