LB 9507 Chapter 5: Raw Data Measurements
5-7
R13
06: MAX. BACKGROUND [RLU/s] 50
(0 = NONE) LAST DISPLAY
Here can you define the limit value for the background. If this
threshold is exceeded, you will be alerted accordingly on the
display and the printout (BKG).
Entering “0” means “no background threshold“.
As soon as you confirm the entry with <enter> you move to the
next display.
R14
06: DELAY LAST INJ./MEAS. [s] 1.0
(0.0 - 300 s) LAST DISPLAY
Enter the delay between the last injection and the start of the
measurement in the range from 0.0 to 300 seconds and press
<enter>.
R15
06: MEASURING TIME [s] 2.0
(0.1 - 200.0) LAST DISPLAY
Enter the desired measuring time (0.1 to 200.0 seconds) for each
tube and press <enter>.
R16
06: NUMBER OF REPLICATES 2
(1 - 10) LAST DISPLAY
Enter the number of replicates and press <enter>. You can define
a maximum of 10 sample tubes as replicates whose measured
values are averaged. “1 means 1 tube, i.e. no replicates.
R17
06:KINETICS PRINTOUT DATA YES
YES NO LAST DISPLAY
The measurement results can be printed in the form of kinetics
data on the built-in thermal printer.