Chapter 5: Raw Data Measurements LB 9507
5-12
R23
06: PROTOCOL ATP-10SEC-1INJ ?
YES NO
You are prompted to confirm that this is the required protocol. If
you select <NO> the program returns to display R22. <YES>
leads to:
R24
COMMENT
You can enter a comment on the following measurement using
the alphanumeric keyboard. Confirm it with <enter>. If you do not
want to enter a comment, just press <enter>.
R28a
SAMPLE 1 TUBE # 1
INSERT TUBE...
Load the tube containing the first sample as defined in the
protocol.
R29
SAMPLE 1 TUBE # 1
START
Press the button beneath the <START> function or the <start>
button to the right of the rotor. Both buttons do the same thing.
The tube will move back into the measurement chamber, and the
programmed measurement procedure will begin:
R30
BACKGROUND MEASURING...
Background measurement is performed, assuming you specified
it in the protocol.