Chapter 5: Raw Data Measurements LB 9507
5-14
Kinetics curve (without data) selected without replicates:
Up to 2000 measurement points per sample displayed as a
curve (according to the measurement time)
Peak maximum and total integral of each sample
Kinetics curve (with data) with replicates:
20 to 50 measurement points per sample
Up to 2000 measurement points per sample displayed as a
curve
Peak maximum and total integral of each sample
Average for replicate group and coefficient of variation
When you remove the sample tube, the measured value display
disappears and the following is requested:
R36
SAMPLE 2 TUBE # 2
INSERT TUBE ...
Load the next sample tube.
R37
SAMPLE 2 TUBE # 2
START
Press the <Start> button. The 2nd tube is measured. If it is a
replicate, “SAMPLE 1” will remain on the display until all tubes of
this group have been measured, while the “TUBE #” is
incremented continuously.
After pressing the <Start> button, the procedure is the same as
described for steps R29 to R34.