LB 9507 Chapter 6: CUT-OFF-Measurements
6-5
C1
READY
MEASURE PROTOCOL -OTHERS-
Select the <PROTOCOL> menu by pressing the respective
button. Then follows:
C2
PROTOCOL
MASTER CURVE CREATE CHANGE
The <MASTER CURVE> is not relevant to CUT-OFF
measurements.
If you press the button below <CHANGE> you can open an
existing protocol by entering its protocol number, and then edit it.
To create a new protocol, press the button below <CREATE>.
The following display appears:
C3
07: PROTOCOL NAME CUT-10SEC-1INJ
The next free number is displayed for the new protocol (in this
example “07:”). Enter a name for this protocol using the
alphanumeric keyboard. You should use a name which specifies
the protocol type and other important features, so that it is easy to
identify later. For example “CUT-10SEC-1INJ”.
Confirm your entry with <enter>.
C4
07: PROTOCOL TYPE
RAW DATA CUTOFF - OTHERS -
Select <CUTOFF>. Now you have to enter the parameters which
are the same for all protocol types:
C5
07: OPERATOR ID
Enter the user name and press <enter>.