Chapter 6: CUT-OFF-Measurements LB 9507
6-6
If you have no injectors selected (see 3.8.2) the display C15
follows. Otherwise C6 follows.
C6
07: ENTER INJECTOR VOLUME 1 [µl]
LAST DISPLAY
Changing or confirmation of the displayed injector volume.
Displayed are the standard settings for injector 1 entered in the
<INSTRUMENT PARAMETERS> menu. These settings can be
changed for this specific protocol.
Setting "0" means the respective injector is disabled.
Range: 25 µL - 300 µL in 5 µL steps.
If the user enters the injector volume not within the 5 µL step
range, the system rounds up or down. Values out of the range are
adjusted to the limits (25 µL or 300 µL).
Example: When entering 43 µL the system rounds up to 45 µL,
when entering 102 µL the system rounds down to 100 µL.
Press <enter> to use the response displayed in the upper-right
corner of the LCD.
C7
07: ENTER INJECTOR VOLUME 2 [µl]
LAST DISPLAY
Changing or confirmation of the displayed injector volume.
Displayed are the standard settings for injector 2 entered in the
<INSTRUMENT PARAMETERS> menu. These settings can be
changed for this specific protocol.
Setting "0" means the respective injector is disabled.
<LAST DISPLAY> takes you back to the previous display.
Setting "0" takes you to display C10.
Setting >0 takes you to display C8.
Press <enter> to use the response displayed in the upper-right
corner of the LCD.
C8
07: SEQUENCE OF INJECTIONS 1->2
1->2 2->1 LAST DISPLAY
If you have selected 2 injectors, you have to define the sequence
of the injections here: injector 1 first and then injector 2 or vice
versa. Make your selection by pressing the button below the