Chapter 6: CUT-OFF-Measurements LB 9507
6-8
C13
07: MAX. BACKGROUND [RLU/s] 50
(0 = NONE) LAST DISPLAY
Here can you define the limit value for the background. If this
threshold is exceeded, a direction is displayed and printed out.
Entering “0” means “no background threshold“. As soon as you
confirm the entry with <enter> you move to the next display.
C14
07: DELAY LAST INJ./MEAS. [s] 1.0
(0.0 - 300) LAST DISPLAY
Enter the delay between the last injection and the start of the
measurement in a range from 0.0 to 300 seconds and press
<enter>.
C15
07: MEASURING TIME [s] 2.0
(0.1 - 200.0) LAST DISPLAY
Enter the desired measuring time (0.1 to 200.0 seconds) for each
tube and press <enter>.
C16
07: REPL. OF NEG. CTRL. (0-10) 3
LAST DISPLAY
Enter the number of replicates of the negative control and press
<enter>. You can define a maximum of 10 replicates whose
measured values are averaged.
C17
07: NUMBER OF POS. CTRL (0-3) 2
LAST DISPLAY
Enter the number of positive controls (0 to 3 allowed). If you enter
0, the displays C18 and C19 are skipped.