EasyManua.ls Logo

LUMAT LB 9507 - Page 85

LUMAT LB 9507
222 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
LB 9507 Chapter 6: CUT-OFF-Measurements
6-9
C18
07: REPL. OF POS. CTRL.1 (1-10) 2
LAST DISPLAY
Enter the number of replicates of the first positive control (up to 10
allowed).
C19
07: REPL. OF POS. CTRL.2 (1-10) 2
LAST DISPLAY
Enter the number of replicates of the second positive control (up
to 10 allowed).
C20
07: REPL. OF SAMPLES (1-10) 2
LAST DISPLAY
Enter the replicate number for all patient samples (1-10).
C21
07: LOW CUTOFF OFFSET 120
LAST DISPLAY
C22
07: LOW CUTOFF FACTOR 1.0
LAST DISPLAY
C23
07: LOW CUTOFF LPOS FACT. 0.0
LAST DISPLAY
These values are used to calculate the low CUT-OFF values:
Low Cut-off = Offset + mean value (NEG) * Low Cut-off factor +
mean value (1st POS.CTRL.) * LPOS. FACTOR (see section 6.1).
C24
07: HIGH CUTOFF OFFSET 120
LAST DISPLAY

Table of Contents