Chapter 6: CUT-OFF-Measurements LB 9507
6-10
C25
07: HIGH CUTOFF FACTOR 0
LAST DISPLAY
C26
07: HIGH CUTOFF LPOS FACT. 1
LAST DISPLAY
The formula for the HIGH CUT-OFF threshold is similar to this
one.
C27
07: %CV FLAG FOR NEG. CTRL. 15
LAST DISPLAY
The negative control mean value is labelled “HICV” in the result
printout if its coefficient of variation is higher than the value
entered here.
C28
07: %CV FLAG FOR POS. CTRL. 20
LAST DISPLAY
Positive standard mean values are labelled “HICV” if their
coefficients of variation are higher than the value entered here.
C29
07: %CV FLAG FOR SAMPLES
LAST DISPLAY
If this CV limit is exceeded by patient samples, a flag appears in
the printout.
C30
07: LOW RLU FLAG NEG. CTRL. 200
LAST DISPLAY