LB 9507 Chapter 6: CUT-OFF-Measurements
6-11
C31
07: HIGH RLU FLAG NEG. CTRL. 600
LAST DISPLAY
Lower and upper limit for marking negative controls: a FLAG
appears in the printout when the value is not within the specified
value range.
“LOW” if measured value < LOW RLU FLAG
“HIGH” if measured value > HIGH RLU FLAG
C32
07: INPUT ALL CORRECT ? YES
YES NO LAST DISPLAY
This display appears at the end of the input sequence. If you
press the button below <YES> the entered parameters are stored
and the program returns to the main menu (C1). <NO> takes you
back to the 1st display of this protocol (C5).
Note:
The protocol is not stored if you exit this protocol sequence via
<exit>!