LB 9507 Chapter 6: CUT-OFF-Measurements
6-15
C48
REMOVE TUBE...
Remove the tube from the sample holder if there is one. This
display appears if there is a tube in the sample holder. So, if a
tube already is in the correct position before you start
measurement, you will have to lift the tube briefly. You will then
see:
C48a
NEG. CTRL REPL.#1TUBE#1
INSERT TUBE ...
Load the tube containing the first negative control (as defined in
the protocol).
C49
NEG. CTRL REPL.#1TUBE#1
START
Press the <start> button. The tube will move back into the
measurement chamber, and the programmed measurement
procedure will begin:
C50
BACKGROUND MEASURING...
REMOVE TUBE...
A background measurement is performed, provided you have
defined one in the protocol.
C51
BACKGROUND TOO HIGH !
This display only appears when the background threshold you
have defined in the protocol is exceeded. If this happens the
printout of measurement values is flagged BKG.The steps which
the instruments goes through are displayed and the measurement