Chapter 6: CUT-OFF-Measurements LB 9507
6-16
parameters from the active protocol are printed out at the same
time.
C51
INJECTING ...
C52
MEASURING ...
C53
SAMPLE-RLU 133 REMOVE
BACKGR. (RLU/S) 30.0 TUBE
Once measurement is complete, the first line shows the result of
the measurement in RLUs, the second line (if defined in the
protocol) shows the result of the background measurement.
The results are printed in 4 columns:
- Measured RLU; for replicates this represents the mean of the
replicate group involved
- Coefficient of variation of the replicate group
-Flag
- Result: indication of the range
The results are classified according to negative controls, positive
controls and samples.
After removing the sample tube, the measured value display is
cleared. You are prompted to:
C54
NEG. CTRL. REPL. # 2 TUBE # 2
INSERT TUBE ...
Load the next tube containing the sample defined on the first line
(in this case the 2nd negative control). The start request will then
appear on the display: