LB 9507 Chapter 6: CUT-OFF-Measurements
6-17
C55
NEG. CTRL. REPL. # 2 TUBE # 2
START
Start the measurement by pressing the <start> button.
In accordance with the selected protocol, you are asked for the
defined number of replicates for negative and then positive
controls. Once a replicate group has been measured, the mean
value and the coefficient of variation for the group are given.
C56
ENTER REPL. NO. TO DISCARD
NONE
If you want to delete a measured replicate, enter the number of
the respective control that is indicated on the result printout.
Otherwise press the button below <NONE>.
The printout shows when a replicate has been deleted. The
deleted replicate is not removed from the memory. When you
load the measured data again (<OPERATOR FUNCTIONS:
DATA EVALUATION: STORED DATA>, see chapter 11.4), the
original measurement data will be used.
C57
CHANGE NEG. CTRL ?
YES NO
<YES>: Here you can edit the mean value. A correction is noted
on the printout. The change is not stored to the memory, so that
the original mean value will be displayed when calling the stored
measurement data (<OPERATOR FUNCTIONS: DATA
EVALUATION: STORED DATA>, see chapter 11.4).
<NO>: Continue the measurement.
Following mean value calculation and a possible correction, the
evaluation of the controls is output to the printer.
Then, the patient samples are requested and mean value (RLU),
variation coefficient, flag and range are printed out for each
sample and each replicate group.