EasyManuals Logo

METREL MI 2893 User Manual

METREL MI 2893
241 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #184 background imageLoading...
Page #184 background image
MI 2893 / MI 2892 / MI 2885 Measurement methods
184
Fundamental phase power measurements
Standard compliance: IEEE STD 1459-2010
All fundamental power measurements are calculated from fundamental voltages and currents obtained
from harmonic analysis (see section 5.1.8 for details).
Fundamental phase active power:
pp
IUfundPfundPfundP
IUP
cos
[W], p: 1,2,3
(17)
Fundamental apparent and reactive power and power factor are calculated according to the following
equations:
Fundamental phase apparent power:
fundPfundPfundP
IUS
[VA], p: 1,2,3
(18)
Fundamental phase reactive power:
pp
IUfundPfundPfundP
IUQ
sin
[var], p: 1,2,3
(19)
Phase displacement power factor:
p
p
pp
S
P
DPF
cos
, p: 1,2,3
(20)
Positive sequence (total) fundamental power measurements
Standard compliance: IEEE STD 1459-2010
According to the IEEE STD 1459, positive sequence power (P
+
, Q
+
, S
+
) are recognised as very important
intrinsic power measurements. They are calculated according to the following equation:
Positive sequence active power:
cos3 IUP
tot
[W],
(21)
Positive sequence reactive power:
sin3 IUQ
tot
[var],
(22)
270
0
+DPFind
Lead
Lag
+DPFcap
-DPFcap
-DPFind
180
0
90
0
0
0
+P-P
-Q
+Q
I
II
III IV

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the METREL MI 2893 and is the answer not in the manual?

METREL MI 2893 Specifications

General IconGeneral
BrandMETREL
ModelMI 2893
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals