MI 2893 / MI 2892 / MI 2885 Measurement methods
189
Total displacement power factor (vector):
fundtot
fundtot
tot
Sv
P
DPFv
All fundamental power measurements are calculated from fundamental voltages and currents obtained
from harmonic analysis (see section 5.1.8 for details).
Total Arithmetic fundamental power measurements
Standard compliance: IEC 61557-12 Annex A and IEEE STD 1459-2010 Section 3.2.2.5
Total Arithmetic fundamental active, reactive and apparent power and total displacement arithmetic
power factor are calculated according to the following equation:
Total fundamental active power:
321 fundfundfund
fundtot PPPP
Total apparent power (arithmetic):
[VA],
Total nonactive power (arithmetic):
22
fundtotfundtotfundtot
PSaQa
Total power factor (arithmetic):
fundtot
fundtot
tot
Sa
P
DPFa
All fundamental power measurements are calculated from fundamental voltages and currents obtained
from harmonic analysis (see section 5.1.8 for details).
5.1.7 Energy
Standard compliance: IEC 62053-21 Class 1S, IEC 62053-23 Class 2 ... MI 2893/MI 2892
Standard compliance: IEC 62053-21 Class 1, IEC 62053-23 Class 2 ... MI 2885
Energy measurement is divided in two sections: ACTIVE energy based on active power measurement
and REACTIVE energy, based on fundamental reactive power measurement. Each of them has two
energy counters for consumed and generated energy.
Calculations are shown below:
[kWh], p: 1,2,3, tot
Generated:
Reactive energy:
Consumed:
m
i
pCap
m
i
Iindp
iTiQiTiQEq
11
)()()()(
[kvarh], p: 1,2,3, tot
Generated:
m
i
pInd
m
i
pCapp
iTiQiTiQEq
11
)()()()(