MI 2893 / MI 2892 / MI 2885 Measurement methods
183
Combined phase active power:
N
j
j
p
j
p
N
j
j
pp
IU
N
p
N
P
11
11
Combined apparent and nonactive power, and power factor are calculated according to the following
equations:
Combined phase apparent power:
Combined phase nonactive power:
Total combined power measurements
Standard compliance: IEEE STD 1459-2010
Total combined (fundamental + nonfundamental) active, nonactive and apparent power and total power
factor are calculated according to the following equation:
Total apparent power (effective):
Total power factor (effective):
In this formula U
e
and I
e
are calculated differently for three phase four wire (4W) and three phase three
wire (3W) systems.
Effective voltage Ue and current Ie in 4W systems:
18
)(3
2
31
2
23
2
12
2
3
2
2
2
1
UUUUUU
Ue
Effective voltage U
e
and current I
e
in 3W systems: