A5/A3™ Operating Instructions 046-003777-00 4 - 13
Preoperative Tests Leak and Compliance Tests
MACHINE NON-FUNCTIONAL
Automatic Circuit Leakage: Pass
Compliance Test: XX.X mL/cmH
2
O
Safety Valve Control: Fail
Example:
Safety valve control test or pressure verification test failed.
Select Retry to repeat the Automatic Circuit Leak Test &
Compliance Test.
or
Contact service if this error condition persists.
NOTE: The Service Access
button is only available
to Mindray-authorized
service personnel and
requires a service
password.
Time out
Example:
Test result cannot be shown due to an internal
communication error.
Select Retry to repeat the Automatic Circuit Leak Test &
Compliance Test.
or
Select Override to skip the test.
RESULTS COMMENTS/OPTIONS
TABLE 4-3 Automatic Circuit Leak and Compliance Test Results