Make AG Settings AG and O2 Concentration Monitoring (Optional)
8 - 6 046-003777-00 A5/A3™ Operating Instructions
8.5 Make AG Settings
Perform the settings below when the anesthesia system is configured with an external AG
module.
8.5.1 Set CO
2
Unit
To change the CO
2
Unit:
1. Select Setup softkey > System tab.
2. Select the CO
2
Unit button.
3. Choose between mmHg, kPa and %.
4. Select the Accept button to confirm the change.
8.5.2 Set CO
2
Placement
To change the CO
2
Placement:
1. Select Setup softkey > Display tab.
2. Select the CO
2
Placement button.
3. Choose between TOP and Bottom.
4. Select the Accept button to confirm the change.
8.5.3 Set CO
2
Scale
To change the CO
2
Scale:
1. Select Setup softkey > Display tab.
2. Select the CO
2
Scale button.
3. Choose between 0-40, 0-60 and 0-80.
4. Select the Accept button to confirm the change.
8.5.4 Gas Bench Flow Rate
To change the Gas Bench Flow Rate:
1. Select Setup softkey > General tab.
2. Select the Gas Bench Flow Rate button.
3. Choose between High (recommended), Med and Low, as follows:
High: 200 ml/min for high volume watertrap; 120 ml/min for low volume watertrap
Med: 150 ml/min for high volume watertrap; 90 ml/min for low volume watertrap
Low: 120 ml/min for high volume watertrap; 70 ml/min for low volume watertrap
4. Select the Accept button to confirm the change.