EasyManua.ls Logo

Mindray A5 - Demographics; Ventilation Modes

Mindray A5
310 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
A5/A3™ Operating Instructions 046-003777-00 C - 9
Parameters and Factory Defaults Demographics
C.6 Demographics
C.7 Ventilation Modes
* SIMV-PC available only on A5.
PARAMETER RANGE DEFAULT
Patient ID
Bed
First Name
Room
Last Name
Point of Care
DOB
Age
Weight (Lbs.)
Facility
OBJECT RANGE DEFAULT
Current selection saved
when powered off
Ventilation
Mode Tab
VCV, SIMV-VC, PCV,
SIMV-PC*, PS
VCV Yes
VENTILATION MODE PARAMETERS
Manual Bypass**, Alarms
VCV Vt, Rate, I:E, Tpause, PEEP, Plimit
SIMV-VC Vt, Rate, Tinsp, Tpause, PEEP, Plimit, PS (On/Off), ΔP, Tr i g g er, Ts l o p e
PCV VtG**, PlimVG**, Pinsp, Rate, I:E, PEEP, Tslope
SIMV-PC** Pinsp, Rate, Tinsp, PS (On/Off), ΔP, Tr i gg e r, P E E P, Ts l o p e
PS Min Rate, ΔP, Trigger, PEEP, Tslope, Apnea Ti
** SIMV-PC, VtG, PlimVG, and Bypass are available only on A5.