A5/A3™ Operating Instructions 046-003777-00 7 - 25
Maintenance Cleaning and Disinfection
7.10.11 Breathing System Block
1. Remove all of the following components from the breathing system block:
Bellows Assembly
•Oxygen Sensor
Inspiratory and Expiratory Valves (all components)
•APL Valve
•PAW Gauge
Bag Arm
Absorber Canister
Inspiratory and Expiratory Flow Sensors
2. Remove the absorber canister (see section7.10.10 (page 7-22) "Absorber Canister").
3. Press and hold the buckle on the bypass assembly to take out the bypass assembly
downward.
FIGURE 7-22 Remove Bypass Assembly
4. Pull out the canister bottom plate upward.
FIGURE 7-23 Remove Canister Bottom Plate
Buckle
Bypass Assembly
Canister Bottom Plate