Monitor Module Product Specifications
9 - 18 046-003777-00 A5/A3â„¢ Operating Instructions
9.10.3 Effect of Interfering Gas on AG Measured Value
Gas Contaminants
Quantitive effect(%
ABS
)
3)
CO
2
N
2
O
Agents
1)
O
2
CO
2
0 0.1 0.1 0.2
N
2
O 0.1 0 0.1 0.2
Agents
1) 2)
0 0.1
0.1
4)
1.0
<100 % Xenon 0.1 0 0 0.5
<50 % Helium 0.1 0 0 0.5
<0.1 % Ethanol 0 0 0 0.5
<1 % Acetone 0.1 0.1 0 0.5
<1 % Methane 0.1 0.1 0 0.5
Saturated Isopropanol vapor 0.1 0 0 0.5
Metered dose inhaler propellants Unspecified Unspecified Unspecified 0.5
1) Agents represents one of DES, ISO, ENF, SEV, and HAL.
2) Multiple agent interference on CO
2
, N
2
O and O
2
is typically the same as single agent interference.
3) For CO
2
, N
2
O and Agents, maximum interference from each gas at concentrations within specified accuracy
ranges for each gas. The total interference of all gases is never larger than 5 %
REL
.
4) Interference for one of the five agents with secondary agent.
TABLE 9-23 Effect of Interfering Gas on AG Measured Value

Table of Contents

Other manuals for Mindray A5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A5 and is the answer not in the manual?

Mindray A5 Specifications

General IconGeneral
BrandMindray
ModelA5
CategoryMedical Equipment
LanguageEnglish

Related product manuals