EasyManuals Logo
Home>Mindray>Medical Equipment>A5

Mindray A5 Operating Instructions

Mindray A5
310 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #277 background imageLoading...
Page #277 background image
A5/A3â„¢ Operating Instructions 046-003777-00 C - 5
Parameters and Factory Defaults Setup Menu
C.3 Setup Menu
PARAMETER RANGE DEFAULT
Current selection
saved when powered
off
General Tab: Breathing
System
Warmer On, Warmer
Off
Warmer On No
General Tab: Gas Bench
Flow Rate
Adult watertrap:
Low (120 ml/min),
Med (150ml/min),
High (200 ml/min)
Infant watertrap:
Low (70 ml/min),
Med (90 ml/min),
High (120 ml/min)
Low (120 ml/min) Yes
Display Tab: Pressure
Display
Mean, PLAT PLAT Yes
Display Tab: Plimit Line On/Off On Yes
Display Tab: Screen
Brightness
level 1-10 5 Ye s
Display Tab: Key Click
Volume
level 1-10 3 Ye s
Display Tab: CO
2
Placement
Top, Bottom Top Yes
Display Tab: Gas Scales:
CO
2
Scale
0-40 mmHg,
0-60 mmHg,
0-80 mmHg
0-60 mmHg Yes
Display Tab: Gas Scales:
Des Scale
0-6.0%,
0-9.0%,
0-18.0%
0-9.0% Yes
Display Tab: Gas Scales:
Sev Scale
0-2.0%,
0-4.0%,
0-8.0%
0-4.0% Yes
Display Tab: Gas Scales:
Iso Scale
0-1.2%,
0-2.5%,
0-5.0%
0-2.5% Yes
Display Tab: Gas Scales:
Hal Scale
0-1.2%,
0-2.5%,
0-5.0%
0-2.5% Yes
Display Tab: Gas Scales:
Enf Scale
0-1.2%,
0-2.5%,
0-5.0%
0-2.5% Yes
Display Tab: Gas Scales:
O
2
Scale
0-35%,
0-50%,
0-100%
0-100% Yes
Display Tab: Gas Scales:
N
2
O Scale
0-35%,
0-50%,
0-100%
0-100% Yes

Table of Contents

Other manuals for Mindray A5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A5 and is the answer not in the manual?

Mindray A5 Specifications

General IconGeneral
BrandMindray
ModelA5
CategoryMedical Equipment
LanguageEnglish

Related product manuals