Setup Menu Parameters and Factory Defaults
C - 6 046-003777-00 A5/A3â„¢ Operating Instructions
Display Tab: Waveform
Display
No module:
Volume, Flow
AG module:
Volume, Flow, AA, O
2
,
N
2
O
Flow Yes
System Tab: Language CHINESE,ENGLISH,
FRENCH, SPANISH,
PORTUGUESE,
RUSSIAN, TURKISH,
DUTCH
ENGLISH Yes
System Tab: Default
Settings: Default Patient
Size
Adult, Pediatric, Infant Infant Yes
System Tab: Default
Settings: Default Vent
Mode
VCV, SIMV-VC, PCV,
SIMV-PC, PS
VCV Yes
System Tab: Manage
Defaults
Save as O.R. Defaults,
Load O.R. Defaults,
Restore Partial Defaults,
Import Defaults,
Export Defaults
Save as O.R.
Defaults
No
System Tab: Change
Password
— — Yes
System Tab: Units:
Pressure
cmH
2
O, hPa, mbar cmH
2
O Yes
System Tab: Units: CO
2
mmHg, kPa, % mmHg Yes
System Tab: Clear
Historys
On, Off Off Yes
System Tab: Time
Settings: Daylight
Savings
Manual, Auto Manual Yes
System Tab: Network:
This Machine: Configure
Ethernet: IP Address
0 - 255 192.168.23.250 Yes
System Tab: Network:
This Machine: Configure
Ethernet: Subnet
0 - 255 255.255.255.0 Yes
System Tab: Network:
This Machine: Configure
Ethernet: Default
Gateway
0 - 255 — Yes
System Tab: Network:
This Machine: Configure
Serial: Baud Rate
4800, 9600, 57600,
115200
9600 Yes
System Tab: Network:
This Machine: Configure
Serial: Parity
Odd, Even, None None Yes
System Tab: Network:
This Machine: Configure
Serial: Data Bits
8, 7, 6, 5 8 Yes
PARAMETER RANGE DEFAULT
Current selection
saved when powered
off