Operating Temperature: +10 to +40°C
+50 to 104°F
Storage Temperature: -20 to +60°C
-4 to 140°F
oxygen sensor: -20 to +50°C
Humidity (Operating and Storage): 15 to 90% RH, non-condensing
Atmospheric Pressure (Operating): 70 kPa to 106.7 kPa
Atmospheric Pressure (Storage): 50 kPa to 106.7 kPa
Resistance to Ingress of Water: Complies with the requirements of clause 11.6.3 in IEC
60601-1 and also the requirements in IEC 60529 for
protection against vertically falling water drops (IPX1)