Silencing Alarms Alarms and Messages
6 - 8 046-003777-00 A5/A3™ Operating Instructions
6.4 Silencing Alarms
When an alarm condition occurs and the alarm audio is sounded, the user can select the Silence
softkey at the bottom of screen to silence the alarm audio. In silenced status, all the alarm indicators
work normally except audible alarm tones.
Select Silence softkey to silence all currently sounding alarm tones. The alarm will sound if a new
alarm occurs.
If the silenced alarms contain medium or high level alarms, the alarm audio will be paused for 120
seconds. The alarm silence icon and 120 second countdown time appear at the top of the screen.
Select again to resume the alarm audio.
NOTE: The alarm will sound if that a new alarm occurs while the system is in an
audio-paused state. If this occurs, you can select the Silence softkey
again to silence the new alarm and reset the silence countdown timer to
120 seconds.
FIGURE 6-5 Alarm Audio-paused
Found in software bundle version 02.04.00 and later
If the silenced alarms are only low level alarms, the alarm audio will be turned off till there is a new
alarm occurs.
NOTE: The alarm will sound if that a new alarm occurs while the system is in an
audio-off state. If the new alarm is low level alarm, you can select the
Silence softkey again to turn off the new alarm audio. If the new alarm
is medium or high level alarm, you can select the Silence softkey again
to silence the new alarm for 120 seconds.
FIGURE 6-6 Alarm Audio-off