A5/A3™ Operating Instructions 046-003777-00 5 - 5
Operations Oxygen Sensor Calibration
5.3.2 Select the Patient Size (Adult, Pediatric, Infant)
Patient size can only be changed when the current ventilation mode is Manual mode, Standby
mode or Monitor mode (available with AG module).
1. Select Manual mode or select the End Case button (in the Manual tab) to enter Standby
mode.
2. Select the Patient Size softkey at the top left of the main screen. The softkey displays
Adult”, “Pediatric”, or “Infant”.
3. Select the Patient Size: Adult, Pediatric, or Infant.
4. Select the Accept softkey to finalize your selection.
NOTE: The A5/A3 saves the latest patient parameter settings (VCV, PCV, PCV-
VG (A5 only), PS, SIMV-VC, SIMV-PC (A5 only), and Alarms) for each
patient type: Adult, Pediatric, and Infant. Changing to another patient
type does not erase the parameter settings from the previous patient
type. For example, changing from Adult to Pediatric and back to Adult
will result in the Adult patient parameter settings still saved.
FIGURE 5-5 Patient Size Setup Menu
5.4 Oxygen Sensor Calibration
If oxygen sensor calibration is needed, please see ‘‘O
2
Sensor Calibration on page 7-6.