A5/A3™ Operating Instructions 046-003777-00 5 - 25
Operations Parameter Monitoring (Waveforms)
.
FIGURE 5-28 Example Simulated Sev vs. Time Waveform (available with the AG module)
5.11.5 Waveform Autoscaling
If the measured values of Paw, Flow, or Volume are larger than the boundary at the end of the
breath cycle, the system will autoscale the Paw, Flow, or Volume at the beginning of next breath
cycle.
If the measured values of Paw, Flow, or Volume are less than the boundary minus a margin (see
TABLE5-1) at the end of two continuous breath cycles, the system will autoscale the Paw, Flow, or
Volume at the beginning of the next breath cycle.
SCALE MARGIN
Paw 10 cmH
2
O if PAW ≥ 30 cmH
2
O
3 cmH
2
O if PAW < 30 cmH
2
O
Flow 10 L/min if Flow ≤ 30 L/min
15 L/min if Flow > 30 L/min
Volume 25 ml if volume ≤ 100 ml
100 ml if volume > 100 ml
TABLE 5-1Autoscaling Margins of Paw, Flow, and Volume