A5/A3™ Operating Instructions 046-003777-00 2 - 7
Installation Install the Vaporizer
2.3.1 Filling and Draining the Vaporizer
Install the vaporizers with a Selectatec interlock system that are compliant to ISO 80601-2-13 on
the A5 unit.
See sectionA.9 (page A-4) "Vaporizers". Refer to the vaporizer manufacturers
Instructions For Use for filling or draining the vaporizer and other information.
WARNING: Ensure that the correct anesthetic agent is used. The vaporizer is
designed with the specific anesthetic agent named on it and further
indicated by color coded labeling. The concentration of the anesthetic
agent actually output will vary if the vaporizer is filled with the wrong
agent.
WARNING: Do not reuse the agent drained from the vaporizer. Treat as a hazardous
chemical and follow local regulations for proper disposal.