A5/A3™ Operating Instructions 046-003777-00 C - 15
Parameters and Factory Defaults Ventilation Parameter Relationships
C.9 Ventilation Parameter Relationships
VENTILATION
MODE Parameter
Parameter
Relationship Equation (s)
VCV Rate
4 ≤ Rate ≤ 100
Vt
20 ≤ Vt ≤ 1500
Plimit Plimit ≥ PEEP+5
10 ≤ Plimit ≤ 100
SIMV-VC Rate
4 ≤ Rate≤ 100
Vt
20 ≤ Vt ≤1500
P P ≤ Plimit-PEEP
3 ≤ P ≤ 50
Plimit Plimit ≥ PEEP+5
Plimit ≥ P+PEEP
10 ≤ Plimit ≤ 100