A5/A3â„¢ Operating Instructions 046-003777-00 7 - 15
Maintenance Cleaning and Disinfection
CAUTION: If moisture remains in the bellows after cleaning, the bellows surface
folds may become tacky and prevent the bellows from properly
expanding. Ensure all moisture is removed from the bellows after
cleaning.
c. After all bellows components are completely dry, inspect them for damage before
disinfection or re-assembly and functional testing.
d. If disinfecting the bellows components, continue with step 7, otherwise skip to step 8.
7. Disinfection
NOTE: Ensure that all bellows components have been cleaned as described in
step
6 before disinfecting.
See section7.10.2 (page 7-11) "Cleaning and Disinfecting Agents / Autoclaving" and use an
approved disinfecting agent for all bellows components while adhering to facility policies and
procedures.
8. Connect the bellows to the breathing system by reassembling all components in the reverse
order. Prior to use after cleaning or disinfecting, power up the system and follow the on-
screen prompts to perform the leak test and the compliance test (
see section4.5 (page 4-9)
"Leak and Compliance Tests").
7.10.5 Inspiration and Expiration Valves
The following procedure is written generically for a single, unspecified valve. It should be performed
on both the inspiration and expiration valves.
FIGURE 7-10 Location of Expiration and Inspiration Valves
1. Remove the valve dome (see FIGURE 7-11), turning it counterclockwise.
Expiration
Valve
Inspiration
Valve

Table of Contents

Other manuals for Mindray A5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A5 and is the answer not in the manual?

Mindray A5 Specifications

General IconGeneral
TypeAnesthesia Machine
BrandMindray
ModelA5
Power Supply100-240 V AC, 50/60 Hz
Ventilation ModesSIMV, PSV
FlowmetersO2, N2O, Air
Vaporizer CapacityUp to 3 vaporizers
Safety FeaturesO2 failure protection, low pressure alarm
Gas ModuleOptional integrated gas module
Integrated Air CompressorOptional
Patient MonitoringIntegrated monitoring with optional modules

Related product manuals