A5/A3™ Operating Instructions 046-003777-00 xvii
Notes Introduction
CAUTION: Turn the flow controls slowly. To avoid damaging the control valves, do
not turn further when the flowmeter reading is outside the range.
When turning a flow control knob clockwise to decrease flow, the
flowmeter should reach zero before the knob reaches its most
clockwise mechanical stop (Off) position. Do not turn any further when
the knob has reached the Off position.
Similarly, when turning a flow control knob counterclockwise to
increase flow from zero, the flowmeter reading should not indicate a
change from zero until the flow control knob is turned approximately
one (1) rotation counterclockwise from the Off position, and only if
permitted according to the gas ratio control system.
Notes
NOTE: Figures in this manual are provided for reference purposes only.
Screens may differ based on the system configuration and selected
parameters.
NOTE: Put the equipment in a location where you can easily see the screen and
access the operating controls.
NOTE: Keep this manual close to the equipment so that it can be obtained
conveniently when needed.
NOTE: The software was developed in compliance with IEC 60601-1. The
possibility of hazards arising from software errors is minimized.
NOTE: This manual describes all features and options. Your equipment may
not have all of them.
NOTE: The A5/A3 is intended to be operated with its integral Breathing
Pressure monitoring in use.
NOTE: The A5/A3 is intended to be operated with its integral Breathing
Pressure limitation devices in use.
NOTE: The A5/A3 is intended to be operated with its integral Exhaled Volume
monitoring in use.
NOTE: The A5/A3 is intended to be operated with its integral Breathing System
integrity Alarm System in use.
NOTE: The A5/A3 is intended to be operated with its integral Continuing
Pressure Alarm in use.
NOTE: The A5/A3 is intended to be operated with its integral O
2
monitoring in
use.
NOTE: The A5/A3 is intended to be operated with an external CO
2
monitor
complying with ISO 80601-2-55. Connection to the CO
2
monitor should
be via a sample line from the patient circuit.
NOTE: The Anesthesia Vapor Delivery Device is to be used with an Anesthetic
Agent Monitor complying with ISO 80601-2-55. Connection to the
Agent monitor should be via a sample line from the Patient Circuit.
NOTE: Continuously monitor the anesthetic agent concentration when using
the Anesthesia System to ensure accurate output of the anesthetic
agent.