EasyManua.ls Logo

Mindray A5 - Sensor Calibration

Mindray A5
310 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
O
2
Sensor Calibration Maintenance
7 - 6 046-003777-00 A5/A3™ Operating Instructions
7.6 O
2
Sensor Calibration
Perform O
2
calibration when the measured value of O
2
concentration has a large deviation from
other reference sources or when the O
2
sensor is replaced. If the O
2
sensor is replaced, 100 % O
2
sensor calibration is required.
For continued O
2
sensor accuracy, the A5/A3 checks for 21 % O
2
calibration approximately every 72
hours. The A5/A3 prompts the user for 21 % O
2
calibration as follows:
When the machine is powered on, if more than 72 hours have elapsed since the last
successful calibration, the prompt message "Calibrate O
2
sensor for 21 %" is displayed. The
message disappears after successful calibration.
If the machine is kept powered on, the prompt message Calibrate O
2
sensor for 21 %” is
displayed at the next Standby mode after 5 AM after 72 hours have elapsed since the last
successful calibration.
NOTE: If the alarm message "RT Clock Needs Battery" or "RT Clock Failure" is
displayed, the prompt message "Calibrate O
2
sensor for 21 %” is
disabled.
The O
2
sensor must be removed from the breathing system before calibrating it at 21 %. The O
2
sensor can be reinstalled after verifying that there is no water build-up in the O
2
sensor and its
installation part.

Table of Contents

Other manuals for Mindray A5

Related product manuals