Cleaning and Disinfection Maintenance
7 - 28 046-003777-00 A5/A3â„¢ Operating Instructions
7.10.12 Active Anesthetic Gas Scavenging System
7.10.12.1 AGSS and Waste Gas Transfer Hose
1. Disconnect the EVAC hose from the AGSS (see FIGURE 7-26).
2. Remove the AGSS and Transfer Hose from the A5/A3.
FIGURE 7-26 AGSS and Transfer Hose Removal
3. Clean the outer surface of the AGSS and Transfer Hose with a soft, lint-free cloth and a
recommended cleaning agent (
see section7.10.2 (page 7-11) "Cleaning and Disinfecting
Agents / Autoclaving"). Allow to dry thoroughly.
4. Remove the top of the AGSS (see FIGURE7-27). Inspect the AGSS filter and shake it over a
waste container to clean it as necessary. If the filter must be replaced, dispose of the old filter
per local disposal regulations.
EVAC Hose Port
Transfer Hose