Alarm and Prompt Messages Alarms and Messages
6 - 26 046-003777-00 A5/A3™ Operating Instructions
6.6.2 Prompt Messages
6.6.2.1 Prompt Messages Displayed in Alarm Area
MESSAGE TIMEOUT REMARK
Pressure, Volume and
Apnea Alarms are OFF
Correspond status
does not exist.
This Alarms Off icon and message appear on a
white background when the Alarms button in
the Manual mode tab is set to Off.
CO
2
and CO
2
Apnea Alarms
are OFF (software bundle
version 02.04.00 and later)
Correspond status
does not exist.
This message appears when the CO
2
Alarms
button in the Manual mode tab is set to OFF.
Load Configuration Failure 10 sec This message appears when the download or
latest configuration update failed.
DEMO Mode - Not for
Clinical Use
Never This message appears when the system is set to
demo mode from the Service tab.
Service Mode - Not for
Clinical Use
Never This message appears when the machine is
worked in Service mode.
Apnea Ventilation Correspond status
does not exist.
This message appears when the Min Rate
triggers a breath in PS ventilation mode.
Calibrate O
2
sensor for 21 % When the machine is powered on, if more than 72 hours have elapsed
since the last successful calibration, the prompt message "Calibrate O
2
sensor for 21%" is displayed. The message disappears after successful
calibration.
• If the machine is kept powered on, the prompt message “Calibrate O
2
sensor for 21%” is displayed at the next Standby mode after 5am after
72 hours have elapsed since the last successful calibration.
• If the alarm message "RT Clock Needs Battery" or "RT Clock Failure"
is displayed, the prompt message "Calibrate O
2
sensor for 21%" is
disabled.
• If the calibrate time is empty, the prompt message "Calibrate O
2
sensor for 21%" is displayed.
Auto-zero in process Correspond status
does not exist.
This message appears when auto-zeroing of the
pressure sensors is in process.
Fresh Gas Is On Correspond status
does not exist.
This message is displayed if the fresh gas flow
value is flashing in Standby mode.
New functions activated,
please restart!
After the machine
restart
This message appears when activation
successfully completed.
Calibrate O
2
sensor for
100%
It will not disappear
until O
2
sensor
calibration
successfully
performed
This message displays when the 100 %
calibration data could not be revised correctly
after 21% O
2
sensor calibrate successfully.
Could not locate time server It will not disappear
until the Interval of
SNTP Protocol is set
to Off or the time
server is available
again.
This message displays when the Interval SNTP
Protocol is not Off and the time server is
unavailable for five (5) intervals.
External AG Loaded
Successfully
5 s “External AG Loaded Successfully. and “External
AG Unloaded. will not display at the same time.
Only the latest one will be shown when both of
them exist.
External AG Unloaded
Successfully
5 s
TABLE 6-10 Prompt Messages Displayed in Alarm Area