A5/A3™ Operating Instructions 046-003777-00 6 - 27
Alarms and Messages Alarm and Prompt Messages
6.6.2.2 Prompt Messages Displayed in Pop-up Area
External AG Startup / External AG module is starting up. This prompt
message is triggered by External AG.
External AG Warmup / External AG module is warming up. This prompt
message is triggered by External AG.
External AG Zeroing / Gas concentrations cannot be measured during
zero, instead the last measured concentrations
are reported to the application.
Leak Test Not Performed Only appears in
standby mode. It
will disappear
when both the
automatic and
manual leak test
has been
performed.
This message displays when either the
automatic leak test or manual leak test was
skipped from startup, or when the last time that
the leak test was performed was more than 24
hours ago (software bundle version 02.06.00
and later).
MESSAGE TIMEOUT REMARK
TABLE 6-10 Prompt Messages Displayed in Alarm Area
MESSAGE TIMEOUT REMARK
Can only End Case in
Manual Mode!
5 sec This message appears if the End Case
button is selected when the Manual
switch is set to Auto and the machine is in
non-standby.
Invalid Age! Please check
DOB or current system
time.
5 sec This message appears after entering the
patient’s date of birth if the calculated age
of the patient is outside the accepted
range (0-150).
Patient Size can only be
changed in Manual Mode
or in Standby
5 sec This message appears when the Patient
Size selection is pressed while the system
is in Automatic Ventilation mode.
Vent modes can only be
changed using "Set Mode"
button below
5 sec This message appears when the Current
Mode area is pressed.
Out of Range 5 sec This message appears when the entered
value is outside the allowable range.
Invalid Password 5 sec This message appears when the entered
password is wrong.
Saving User Configuration
has failed.
5 sec This message appears when the Saving
User Configuration process has failed.
New password input is
inconsistent.
5 sec This message appears when the new
password and the confirmed new
password do not match.
Automatic ventilation
disabled. Check lever on
breathing system.
15 sec This message appears when exiting
Standby mode while the Auto/Manual
switch is in Auto position and system is in
the Automatic Ventilation disabled state.
TABLE 6-11 Prompt Messages Displayed in Pop-up Area