Alarm and Prompt Messages Alarms and Messages
6 - 28 046-003777-00 A5/A3™ Operating Instructions
Fresh gas flow detected!
Adjust all flowmeters to
zero
After fresh gas flow is
turned off or after exiting
"Manual Circuit Leak
Test" or "Automatic
Circuit Leak Test &
Compliance Test" screen.
This message appears in the first "Manual
Circuit Leak Test" or "Automatic Circuit
Leak Test & Compliance Test" screen when
fresh gas flow is detected.
Access to System settings
only available in Standby
5 sec This message appears when the current
mode is in non-standby and the user tries
to enter the Setup > System menu.
Can not end case while
fresh gas flow is detected!
5 sec This message appears when user tries to
end the case by pressing the disabled End
Case button while fresh gas is on, Auto/
Manual switch is set to Manual, and the
system is in non-standby.
Set Auto/Manual switch to
Manual position before
starting case
1. When triggered by
turning on fresh gas, it
will disappear after fresh
gas flow is turned off or
Auto/Manual switch is
set to Manual;
2. When triggered by
touching the Waveforms/
Spirometry screen, it will
disappear after 5 seconds
or Auto/Manual switch
is set Manual.
When Auto/Manual switch is in Auto
position and system is in Standby mode,
this message will appear in the following
cases:
1. Turning on fresh gas
2. Touching the Waveforms/Spirometry
screen
Set Auto/Manual switch to
Manual position and adjust
all flowmeters to zero.
5 sec This message appears in the first
"Automatic Circuit Leak Test & Compliance
Test" screen when pressing the disabled
Continue button.
Set Auto/Manual switch to
Auto position and adjust
all flowmeters to zero.
5 sec This message appears in the first "Manual
Circuit Leak Test" screen when pressing the
disabled Continue button.
Can only End Case in
Manual Mode!
5 sec This message displays when the Auto/
Manual switch is in Auto position and the
system is in non-Standby, then if the user
presses the disabled End Case button.
MESSAGE TIMEOUT REMARK