A5/A3™ Operating Instructions 046-003777-00 5 - 7
Operations Input Fresh Gas
5.5.2 Set Anesthetic Agent
NOTE: You do not need to perform this operation if inspiratory anesthetic
agent is not used.
NOTE: The A5/A3 anesthesia system can be mounted with vaporizers
corresponding with Halothane, Enflurane, Isoflurane, Sevoflurane and
Desflurane. Only one vaporizer can be opened at a time because of the
interlock system.
5.5.2.1 Select the Desired Anesthetic Agent
1. Determine the anesthetic agent to be used and then fill the vaporizer.
NOTE: Install the vaporizers with a Selectatec interlock system that are
compliant to ISO 80601-2-13 on the A5 unit. Refer to the vaporizer
manufacturer’s Instructions For Use for filling or draining the vaporizer
and other information.
WARNING: Ensure that the correct anesthetic agent is used. The vaporizer is
designed with the specific anesthetic agent named on it and further
indicated by color coded labelling. The concentration of the anesthetic
agent actually output will vary if the vaporizer is filled with the wrong
agent.
2. Mount the vaporizer filled with anesthetic agent onto the A5/A3 Anesthesia System. See
“Install the Vaporizer” on page 2-5.
5.5.2.2 Adjust the Concentration of Anesthetic Agent
Push and turn the concentration control on the vaporizer to set the appropriate concentration of
anesthetic agent. For details about how to use the anesthetic agent, refer to the Vaporizer
Instructions for Use.