EasyManuals Logo
Home>Mindray>Medical Equipment>A5

Mindray A5 Operating Instructions

Mindray A5
310 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #22 background imageLoading...
Page #22 background image
Introduction Notes
xviii 046-003777-00 A5/A3â„¢ Operating Instructions
NOTE: Check the liquid level of the anesthetic agent before and during all
operations. When the liquid level is below the warning line, more
anesthetic agent needs to be added. Refer to the vaporizer Instructions
For Use for filling the vaporizer and other information.
NOTE: The A5/A3 System is designed to be equipped with an anesthetic vapor
delivery device that complies with ISO 80601-2-13.
NOTE: The A5/A3 battery supply is not a user serviceable component. Only an
authorized service representative can replace the battery supply. If the
system is not used for an extended period, contact a service
representative to have the battery supply disconnected. The batteries
may be subject to local regulations regarding disposal. At the end of
the battery life, dispose of the battery supply in accordance with local
regulations.
NOTE: Areas designated for the servicing of oxygen equipment shall be clean,
free of oil and grease, and not used for the repair of other equipment.
NOTE: Opening the cylinder valve quickly may cause unexpected pressure
differentials and create a potential for fire or explosion arising from
oxygen pressure shocks. Open and shut the cylinder valve slowly.
NOTE: Accuracy of the flowrate may be affected by varying inlet pressure,
varying outlet resistance, or varying ambient temperature.
NOTE: The power device, terminal units and pipeline system can be supplied
by one or several different manufacturers.
NOTE: Regional or national regulations that apply to manufacturers of
medical devices can exist.
NOTE: For the method of connecting A5 to external monitor or other devices,
please see Anesthesia Machine Bracket Installation Instructions.
NOTE: The A5 can be equipped with one scavenger system to provide the best
match with the hospital’s waste-gas disposal system. The scavenger
system shall comply with ISO 80601-2-13.
NOTE: The Anesthesia System is compatible with gases (O
2
, N
2
O, and Air) and
anaesthetic agents (Halothane, Enflurane, Isoflurane, Sevoflurane, and
Desflurane).
NOTE: The leakage of AGSS is measured by the method recommended in ISO
80601-2-13.
NOTE: The Anesthesia System is compatible with gases (O2, N2O, and Air) and
anaesthetic agents (Halothane, Enflurane, Isoflurane, Sevoflurane, and
Desflurane).
NOTE: The leakage of AGSS is measured by the method recommended in ISO
80601-2-13.

Table of Contents

Other manuals for Mindray A5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A5 and is the answer not in the manual?

Mindray A5 Specifications

General IconGeneral
BrandMindray
ModelA5
CategoryMedical Equipment
LanguageEnglish

Related product manuals