Alarm Limits Parameters and Factory Defaults
C - 4 046-003777-00 A5/A3™ Operating Instructions
EtEnf Low Off, 0.0 to (High-0.2)
Step: 0.1
0 % Yes
FiEnf High Off, (Low+0.2) to 5.0
Step: 0.1
2 % Yes
FiEnf Low Off, 0.0 to (High-0.2)
Step: 0.1
0 % Yes
EtIso High Off, (Low+0.2) to 5.0
Step: 0.1
3 % Yes
Etlso Low Off, 0.0 to (High-0.2)
Step: 0.1
0 Yes
FiIso High Off, (Low+0.2) to 5.0
Step: 0.1
2 % Yes
Filso Low Off, 0.0 to (High-0.2)
Step: 0.1
0 % Yes
EtSev High Off, (Low+0.2) to 8.0
Step: 0.1
6 % Yes
EtSev Low Off, 0.0 to (High-0.2)
Step: 0.1
0 Yes
FiSev High Off, (Low+0.2) to 8.0
Step: 0
5 % Yes
FiSev Low Off, 0.0 to (High-0.2)
Step: 0.1
0 % Yes
EtDes High Off, (Low+0.2) to 18.0
Step: 0.1
8 % Yes
EtDes Low Off, 0.0 to (High-0.2)
Step: 0.1
0 % Yes
FiDes High Off, (Low+0.2) to 18.0
Step: 0.1
6 % Yes
FiDes Low Off, 0.0 to (High-0.2)
Step: 0
0 % Yes
EtO2 High Off, (Low+0.2) to 100
Step: 1
88 % Ye s
EtO2 Low Off, 10 to (High-2)
Step: 1
Off % Yes
CO2 Apnea
Delay Time
10 sec, 15 sec, 20 sec,
25 sec, 30 sec, 35 sec,
40 sec
30 sec, Yes
PARAMETER RANGE DEFAULT UNIT
Current selection
saved when
powered off

Table of Contents