Alarm and Prompt Messages Alarms and Messages
6 - 16 046-003777-00 A5/A3â„¢ Operating Instructions
Continuous
Airway
Pressure
Paw in the breathing
circuit > sustained airway
pressure alarm limit for 15
seconds
High No N/A * Yes
Negative
Pressure
Paw < -10 cmH
2
O for 1
second.
High No N/A * Yes
EtCO
2
Too High EtCO
2
> high alarm limit
setting
Medium No Yes No
EtCO
2
Too Low EtCO
2
< low alarm limit
setting
Medium No Yes No
FiCO
2
Too High FiCO
2
> high alarm limit
setting
Medium No Yes No
EtN
2
O Too
High
EtN
2
O > high alarm limit
setting
Medium No Yes No
EtN
2
O Too Low EtN
2
O < low alarm limit
setting
Medium No Yes No
FiN
2
O Too High FiN
2
O > high alarm limit
setting
Medium No Yes No
FiN
2
O Too Low FiN
2
O < low alarm limit
setting
Medium No Yes No
EtHAL Too
High
EtHAL > high alarm limit
setting
Medium No Yes No
EtHAL Too Low EtHAL < low alarm limit
setting
Medium No Yes No
FiHAL Too High FiHAL > high alarm limit
setting
Medium No Yes No
FiHAL Too Low FiHAL < low alarm limit
setting
Medium No Yes No
EtENF Too High EtENF > high alarm limit
setting
Medium No Yes No
EtENF Too Low EtENF < low alarm limit
setting
Medium No Yes No
FiENF Too High FiENF > high alarm limit
setting
Medium No Yes No
FiENF Too Low FiENF < low alarm limit
setting
Medium No Yes No
EtISO Too High EtISO > high alarm limit
setting
Medium No Yes No
EtISO Too Low EtISO < low alarm limit
setting
Medium No Yes No
FiISO Too High FiISO > high alarm limit
setting
Medium No Yes No
FiISO Too Low FiISO < low alarm limit
setting
Medium No Yes No
EtSEV Too High EtSEV > high alarm limit
setting
Medium No Yes No
MESSAGE CAUSE
ALARM
PRIORITY
DISABLED
WHEN
ALARM IS
OFF
DISABLED
IN
STANDBY
MODE
DISABLED
IN
MONITOR
MODE
* N/A - Not Applicable. This alarm message does not exist within this mode and therefore cannot be disabled
or enabled.
TABLE 6-3 Physiological Alarm Messages