8-20
Mortara algorithm
Arrh. event Range Default Step Unit
PVCs High 1 to 100 10 1 /min
Asys. Delay 2 to 10 5 1 s
Vtac Rate 100 to 200 130 5 bpm
Vtac PVC 3 to 12 6 1 beats
Multif. PVC 3 to 31 15 1 beats
Tachy High Adult: 100 to 300
Pediatric: 160 to 300
Adult: 100
Pediatric: 160
5 bpm
Brady Low Adult: 15 to 60
Pediatric: 15 to 80
Adult: 60
Pediatric: 80
5 bpm
8.8.4 Setting the Extended Arrh. (For Mindray Algorithm Only)
The following arrhythmia events are defined as extended arrhythmia:
Extreme Tachy
Extreme Brady
Vent. Brady
Nonsus. Vtac
Multif. PVC
Irr. Rhythm
Pause
Afib
You can select [Main Menu][Maintenance >>][User Maintenance >>]enter the required passwordselect
[Alarm Setup >>], and set [Extended Arrh.] to [Enable] or [Disable]. When [Extended Arrh.] is set to [Disable], the
patient monitor does not analysis the extended arrhythmia events and corresponding alarms are not given.
8.8.5 Reviewing Arrhythmia Events
Please refer to the Review chapter.